Purpose
Hydrodissection separates the cortex from the lens capsule using a fluid wave.
Requirements
-
The fluid wave must be forceful enough to break the adhesions between the capsule and the cortex.
-
The volume must be sufficient to achieve separation, without causing the lens to prolapse out of the capsular bag into the anterior chamber.
-
The fluid wave must be directed behind the lens.
-
Excessive pressure must be avoided in the anterior chamber to prevent iris prolapse.
Technique
Position the cannula 1.5–2 mm beneath the capsulorhexis edge. Inject balanced salt solution (BSS) only once the cannula is correctly positioned.
Advancing the cannula sufficiently beneath the capsulorhexis edge directs the fluid behind the lens. This effectively creates a temporary compartment, allowing a fluid wave of sufficient force and volume to develop. If the cannula is positioned too close to the edge, BSS can escape into the anterior chamber, reducing both the volume and the force of the wave.
Ideally, the cannula should gently lift the capsulorhexis edge. Tilting the cannula in this way also tends to depress the posterior lip of the main incision, allowing viscoelastic to escape from the anterior chamber as fluid is injected. This balances inflow and outflow, substantially reducing the risk of iris prolapse.
With a temporal incision, inject at the 6 and 12 o’clock positions. With a superior incision, inject at 9 and 3 o’clock. These positions are ideal for creating a 180° fluid wave with each injection.
As the wave approaches the opposite side, gently depress the lens with the cannula tip. This allows fluid behind the lens to pass around the equator and flow beneath the anterior capsule. The resulting “second wave” breaks the remaining adhesions between the lens cortex and the anterior capsule.
A characteristic reticular pattern beneath the anterior capsule typically appears—a hallmark of successful hydrodissection. This manoeuvre also releases fluid trapped behind the lens, creating space for a second hydrodissection wave from the opposite side. Ideally, a reticular pattern should then be visible beneath the anterior capsule throughout the full 360°.
Troubleshooting
Iris prolapse
The primary cause of iris prolapse is excessive pressure.
Depress the posterior lip of the main incision to maintain a balance between inflow and outflow.
If iris prolapse occurs, relieve the excess pressure first (see the article on iris prolapse):
-
Depress the posterior lip of the main incision to allow fluid or viscoelastic to escape.
-
If necessary, use the bimanual aspiration cannula, without irrigation, to remove viscoelastic from the anterior chamber until the eye softens. Remember to reintroduce viscoelastic before starting phacoemulsification to protect the cornea?l endothelium.
-
Then inject small, forceful pulses of fluid behind the lens while firmly depressing the posterior lip of the incision to facilitate outflow.
A golden ring without a clear fluid wave
If a golden ring appears instead of a clear hydrodissection wave, the fluid has passed between the epinucleus and the nucleus, producing hydrodelineation.
Repositioning the cannula correctly will allow hydrodissection to be performed.
Failure of the lens to rotate
If the lens does not rotate, perform additional hydrodissection. Look particularly for areas beneath the anterior capsule that lack a reticular pattern, indicating undisturbed cortex. These are the areas where hydrodissection is most likely to have been incomplete.
Nucleus prolapse
If the nucleus prolapses out of the capsular bag, it must be repositioned. Do not simply push it straight down, as this presses the nucleus against the anterior capsule and places considerable tension on the zonules.
Instead, first sweep the nucleus towards the side opposite the prolapse, then gently depress it back into the capsular bag.